GE Healthcare · MRExcite2248346224832122494222402164
Intermittently Unable to Download PSDs Due to Network Switch in the System Cabinet.
Symptom
Intermittent failure in Downloading of PSD's and Autoprescan failures. Intermittently unable to download PSD's and intermittently autoprescan fails with autoshim selected.
Error message
2248346 Pulse generation failed. Probable PSD software error. 2248346 MMCN NSP SCP:sct_prep_or_start 417 Pulse generation failed. Probable PSD software error. Pulse generation timed out. It may be necessary to perform a TPS Reset. 2248321 The callback for SCSM state 0x8 was not received in time. The PS server is not responding, the server might be hung. Please perform a TPS reset to recover.-> Error: 2248321 Host: EXCITEMR Process: NSP File: SCP:sct_wdog_exp Line: 1003 Communication Error. The callback for SCSM state 0x8 was not received in time. The PS server is not responding, the server might be hung. Please perform a TPS reset to recover.-> 2249422 An ermes-triggered error was detected in task sct (0x2FB0948) for ermes number 2248321. Additional information in /usr/g/service/log/vxErmes TrgLg-24a4d2 Error: 2249422 Host: EXCITEMR Process: diagCoord File: excLog.c Line: 1427 Task ID: 0x3dcdef0 Time: 2402164 Object: SCP An ermes-triggered error was detected in task sct (0x2FB0948) for ermes number 2248321. Additional information in /usr/g/service/log/vxErmes TrgLg-24a4d2
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.