Intermittently Unable to Download PSDs Due to Network Switch in the System Cabinet.

Symptom

Intermittent failure in Downloading of PSD's and Autoprescan failures. Intermittently unable to download PSD's and intermittently autoprescan fails with autoshim selected.

Error message

2248346 Pulse generation failed. Probable PSD software error. 2248346 MMCN NSP SCP:sct_prep_or_start 417 Pulse generation failed. Probable PSD software error. Pulse generation timed out. It may be necessary to perform a TPS Reset.

2248321 The callback for SCSM state 0x8 was not received in time. The PS server is not responding, the server might be hung. Please perform a TPS reset to recover.-> Error: 2248321 Host: EXCITEMR Process: NSP File: SCP:sct_wdog_exp Line: 1003 Communication Error. The callback for SCSM state 0x8 was not received in time. The PS server is not responding, the server might be hung. Please perform a TPS reset to recover.->

2249422 An ermes-triggered error was detected in task sct (0x2FB0948) for ermes number 2248321. Additional information in /usr/g/service/log/vxErmes TrgLg-24a4d2 Error: 2249422 Host: EXCITEMR Process: diagCoord File: excLog.c Line: 1427 Task ID: 0x3dcdef0 Time: 2402164 Object: SCP An ermes-triggered error was detected in task sct (0x2FB0948) for ermes number 2248321. Additional information in /usr/g/service/log/vxErmes TrgLg-24a4d2

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution