The configuration file ~sdc/app-defaults/network/dcs.cfg is preconfigured for 1 hour DICOM association timeout.
Symptom
Scanner will stop sending images after an hour, then wait 3 minutes, and start sending the exam again from the beginning causing an endless loop. See Long Description PQR 13113964: Customer comments are below. Problem is the scanner will stop sending images after an hour, then wait 3 minutes, and start send the study again from the beginning causing an endless loop. These studies can contain more than 20k images, some series as high as 11k images. Customer: I am investigating an issue with very large research studies getting stuck in the send queue on our MR scanners. It appears that the scanner is closing the association after one hour then starting to resend the study from the beginning. This eventually causes the study to fail sending in its entirety into PACS because eventually it times out while sending duplicates and never gets to the last images in the study. Do you know if it is possible to change the Maximum amount of time to keep a DICOM association open on the scanners?
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.