Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2248349 The file MRconfig.cfg contains a missing or bad value. See Long Description This error will include one of many different data values that can be corrupted. Some examples are: begTravel boreTempHyst boreTempLev1 boreTempLev2 boreTempOffset digiTuningBoard endTravel fieldStrength gradientAmp gradProcessorType isoVectorZ lineFreq mCoilTRdriver PowerMonType realTimeMon RFampType rrfCalibFlag specRFamp spectroRFampType usssType

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR