Pular para o conteúdo principal
GE Healthcare · CT178574178568184518

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 178574 The following controllers STC/ETC/OBC failed to report initialization status in 120 seconds See Long description for error code variations Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ---------------------------------------------------------------------------- Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds

Error: 184518 Write-Read Test of Address 0x00ffae08 Failed See LD for the full error code Error: 184518 Host: STC Process: Control File: AXitest.c 1.14 Line: 435 Function: No System Function Reported Scan: 0/0/0 Type: None/Unkno

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264806G1 STC AXIAL CONTROL BOARD See LD for details Part #: 46-264806G1 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - FE: READ SN T-1345. IF U455 ONAX BOARD IS BEING SHORTED, THECABLE MAY BE PINCHED. MS8365 4/95 HQ INV ONLY: PCN'S 161734 AND 161831 IMPLEMENTED. STOCK WASPURGED. 8/93 MBELL FLS - FE ASSIST: MS6588 MAY 2002 46-264806G1 = AXIAL BD FOR CT1.0 SEC. AXIAL ROTATION. THIS BOARD HAS THE PALS FOR 1.0 SECOND ROTATION. THE T3202BD =1.0 SEC AXIAL BD. THE T3202KD = .8 SEC AXIAL BD. THE ONLY DIFFERENCE BETWEEN THE T3202KD AND T3202BD ARE TWO PALS. THE T3202BD CAN BE UPGRADED TO THE T3202KD BY CHANGING PALS U211 AND U101. DO YOU STILL WANT TO ORDER THIS PART?? 46-264806G1 STC AXIAL CONTROL BOARD
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues