GE Healthcare · CT1821265009149116349096618211650024
Lightspeed 1.X - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 182126 LAN Watchdog TIMEOUT with OBC (uuid 193) See LD for complete error Wed Sep 20 08:39:07 2006 Error: 182126 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 851 Function: No System Function Reported Scan: 50091/8/0 Type: None/Unknown Exception Level: Secondary Time: 08:39:01:815 Log Series: 724 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:491163 Last Response Rcvd:490966 (in Ticks)). Error: 182116 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). See Long Description for Error Mon Dec 6 22:19:40 2004 Error: 182116 Suite: CT01 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Method: No Method Line: 766 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:19:33:885 Log Series: 27 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. Will need a Hardware Reset. Or Tue Sep 23 19:48:54 2003 Error: 182116 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 795 Function: No System Function
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2253794 H1 TRANSCEIVER BUFFER BOARD_SCHLEIFRING used on LS1.X See Long Description for details Part #: 2253794 Part Type: GP CONSUMABLES Part Weight: 0.20 KGM Repairable: NoReturnable: No Hazardous Material: N Description: H1 TRANSCEIVER BOARD, SCHLEIFRING FRU: Yes Status: Orderable Parts Offered as: Current: 2253794 Harvested: Not AvailableExchange: Not AvailableComments: FLS - TRANSCEIVER BOARD HELIOS 1 FORWARD PRODUCTION MS08992 3/1/00 MIN QTY 10 2 PER SYSTEM PCN 199706GELS - ****************************************************** FOR GPO EUROPE ONLY : PLS PLACE PURCHASE ORDER FOR P1 ORDER TO SUPPLIER LUSO 0026195// MFG IS LOCATED IN GERMANY // PDB 00/10/13,TRANSCEIVER BOARD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required