Pular para o conteúdo principal
GE Healthcare · CTHiSpeedVCT178574178568244540

Attention: The Scanning Hardware is not available

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 178574 The following controllers STC/ETC/OBC failed to report initialization status in 120 seconds See Long description for error code variations Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ---------------------------------------------------------------------------- Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 R

Applies to

Models
HiSpeed, VCT
Product
HiSpeed QX/i - GRE Always MDAS 4 QX_P_M4 HiSpeed QX/i - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an MDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. xxx
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.