Pular para o conteúdo principal
GE Healthcare · CTHiSpeed178568244553244554

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 178568 Failed to establish socket connection to STC in 10 seconds See Long Description for full error Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds

Error: 244553 The STC controller has failed to reset See long description for full error Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset .

Error: 244554 The reset failure of the OBC may be due to the reset failure of the STC As reported on CTi 6.x: Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC.

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264806G1 STC AXIAL CONTROL BOARD See LD for details Part #: 46-264806G1 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - FE: READ SN T-1345. IF U455 ONAX BOARD IS BEING SHORTED, THECABLE MAY BE PINCHED. MS8365 4/95 HQ INV ONLY: PCN'S 161734 AND 161831 IMPLEMENTED. STOCK WASPURGED. 8/93 MBELL FLS - FE ASSIST: MS6588 MAY 2002 46-264806G1 = AXIAL BD FOR CT1.0 SEC. AXIAL ROTATION. THIS BOARD HAS THE PALS FOR 1.0 SECOND ROTATION. THE T3202BD =1.0 SEC AXIAL BD. THE T3202KD = .8 SEC AXIAL BD. THE ONLY DIFFERENCE BETWEEN THE T3202KD AND T3202BD ARE TWO PALS. THE T3202BD CAN BE UPGRADED TO THE T3202KD BY CHANGING PALS U211 AND U101. DO YOU STILL WANT TO ORDER THIS PART?? 46-264806G1 STC AXIAL CONTROL BOARD
Area
CT

Full cause & step-by-step resolution