Pular para o conteúdo principal
GE Healthcare · MRExciteDiscovery2224358224938622279722227973

Browser does not list Exams.

Symptom

Browser does not list Exams. Browser List is Suddenly Empty. Browser does not list Exams. Browser list is suddenly empty.

Error message

2224358 Unable to open image file 2224358 lx-mr TIR update_image_header.c 223 Unable to open image file /export/home1/sdc_image_pool/reserved/IMSseSAA a000V5/8407.2.1.36. Errno = 2

2249386 There was an error creating a file for logging an error message. Filename vxExLg-2122 could not be created. 2249386 lx-mr diagCoord excLog.c 590 Task ID: 0xdd9ef0 Time: 8482 Object: SCP There was an error creating a file for logging an error message. Filename vxExLg-2122 could not be created.

9054 Cannot access directory in read/write mode /export/home1/sdc_image_pool/buf_image May 4 Fri 06:41:18 9054 dbrserver init.c 93 Cannot access directory in read/write mode : /export/home1/sdc_image_pool/buf_image

2227972 Wait time to receive response from TIR exceeded 20secs...still waiting

2227973 Received image transfer response from TIR 2227973 Host: lx-mr Process: NSP File: TPS:ed_wait_on_cm Line: 11 Received image transfer response from TIR

Applies to

Models
Excite, Discovery
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution