Scan Pauses with Gradient Errors with Bright Line Artifact Due to Gradient Processor GP.

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2250109 The GP is reporting a Gradient Clock Stop. The GP stopped receiving a gradient clock signal from the systems cabinet.

2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution