GE Healthcare · MRExcite22501092253850
Scan Pauses with Gradient Errors with Bright Line Artifact Due to Gradient Processor GP.
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2250109 The GP is reporting a Gradient Clock Stop. The GP stopped receiving a gradient clock signal from the systems cabinet. 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.