How MM3 Compressor code is created.

Symptom

What do the Magnet Monitor 3 Compressor Status Codes mean in Insite 2. Need to know what the codes in InSite 2 mean. Need to know what the codes in InSite2 mean.

Error message

Magnet Monitor 3 Compressor Code 04, Klixon Error and Compressor is Stopped.

Magnet Monitor 3 Compressor Code 32, Compressor Stopped Due to Low Pressure and No 24 Volts.

Magnet Monitor 3 Compressor Code 00, Normal Operation.

Magnet Monitor 3 Compressor Code 12, Compressor stopped Due to Low Pressure.

Magnet Monitor 3 Compressor Code 31, Compressor Stopped Due to Over Heat and No 24 Volts.

Magnet Monitor 3 Compressor Code 33, Compressor Stopped Due to Overheat, Low Pressure and No 24 Volts.

Magnet Monitor 3 Compressor Code 30, Compressor Stopped No 24 Volts.

Magnet Monitor 3 Compressor Code 18, Compressor Fuse is Tripped and the Compressor is Stopped.

Magnet Monitor 3 Compressor Code 20, Compressor Powered Off or Cable is Disconnected No 24 Volts.

Magnet Monitor 3 Compressor Code 10, Compressor Stopped for Unknown Reason.

Magnet Monitor 3 Compressor Code 11, Compressor Stopped Due to Overheat.

Magnet Monitor 3 Compressor Code 13, Compressor Stopped Due to Overheat and Low Pressure.

Magnet Monitor 3 Compressor Code 08, Compressor Fuse is Tripped.

Applies to

Product
Magnet Monitor 3. Magnet Monitor 3 Magnet Monitor 3
Software
Magnet Monitor 3 Revision 1.56. Magnet Monitor 3 revision 1.56, distributed to sites with Connectivity automatically post FMI 60755 release on 6/23/2008. See Long Description for Fixes and Notes. Magnet Monitor 3 Revision 1.56 Software (formerly known as 1.55). Fixes New magnet type selection page with menus for both compressors. New remote commands to change the compressor selections. New uploaded properties to show the current compressor selections. A fix to the satellite interface so MM3 doesn't lose the sat link if it gets a malformed entitlement message. Changed configuration files so ABM and BOM magnets with DT500 sensors use the correct lookup tables. Disabled alarms actually won't generate alarms. The net mask no longer has to be entered with extra zeros (i.e. - 255.0.0.0 is now good enough instead of 255.000.000.000). The "Show Digital IO" web page shows all compressor types correctly - no more messages saying that Leybolds are inverted. New alarm limits for ABM and BOM for 2nd_stage_shield (is now 10K to 25K, was previously 3.5K to 10K). Changed Master Interval to 720 minutes for LCC and Non-Zero boil off magnets and 360 minutes for HFOs to reduce dispatch volume. NOTES •Software will be downloaded to all communicating sites except satellite modems. •Currently there is no CD that is orderable as a FRU for sites that do not have communication or have a satellite modem so that the software can be manually updated.
Component
2394952 Magnet Monitor 3 Magnet Monitor 3 manufactured by Advantech starting in 2005, part number 2394952. Magnet Monitor 3 manufactured by Advantech starting in 2005, part number 2394952.
Area
MR

Full cause & step-by-step resolution