Pular para o conteúdo principal
GE Healthcare · MRSignaExcite22256932225512222569922256632225550

ERBTEC RF Amplifier fault 84 after reseting the TPS, generic MDS link errors

Symptom

Unable to scan with Erbtec errors

Error message

2225693 MDS link failure. An MDS device failed to respond to IPG. Wed Oct 29 00:14:08 1997 2225693 1 CMR1 OC0 HPC_PORT_ID PMH.m 1.29 spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0252 Transmitted header: 0x10 Transmitted opcode: 0x02

2225512 Dynamic disable and T/R driver fault. A write to the driver board coil register failed to verify

2225699 MDS Link Failure. An MDS device failed to respond to the IPG on a 2nd try.

2225663 ERBTEC RF Amplifier fault 84. Power Monitor Fault.

2225550 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify See Long Description 2225550 Host: LX_MR1 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6300 Multi-coil array driver error.^A A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis.

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues