Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2244655

Scan Pause and Bright Zipper or Band in Images caused by TWIN GRAD SWITCH BOX.

Symptom

MRIQ Gradient Switch Related Errors Pause Scans and Resultant Images Contain a Bright Zipper or Band Artifact Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. This system is a 1.5T Excite II TwinSpeed running 11.0 M4 software with Service Pack 1 loaded. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. There are several error messages logged which seem to point to the GSC (Gradient Switch Controller). Below are copied the error messages logged at the time of one of these scan pauses and a typical image. The fix was to replace the "TWIN GRAD SWITCH BOX" part number 2253466 in the TAC cabinet. This module contains the GSC. Reference SPR # MRIhc16978. Mon Sep 25 10:12:27 2006 2244648 WHMR1 NSP SCP:GP 668 The GSC reported a problem with the Whole switch on the Y axis. If the problem persists, please call a service engineer. Mon Sep 25 10:12:28 2006 2244655 WHMR1 NSP SCP:GP 1307 The GP received an ALARM message from the GSC. Subsequent error messages should reveal the cause of the alarm. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Y axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Z axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 2 WHMR1 NSP SCP:GP 1069 The gradient system on the X axis recovered successfully. There are no faults at the present time. Author Paul Lyman SME John Schumacher

Error message

2244655 The GP received an ALARM message from the GSC. 2244655 WHMR1 NSP SCP:GP 1307 The GP received an ALARM message from the GSC. Subsequent error messages should reveal the cause of the alarm.

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues