Appears the X and Y Coil Drive is swapped Error Due to DQA Phantom Fluid level.

Symptom

Appears the X and Y Coil Drive is swapped Error In DQA Calibration see long description Auto DQA Calibration failed due to error during analysis. "Appears the X and Y Coil Drive is swapped". The images were correct geometrically. However the analysis reported detecting two objects. The top structure the program was measuring had a dip in it due to low fluid in the Phantom

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
2321556 1.5T DQA Phantom with Loader From the Signa HD & HDx 1.5T FRU’S DIRECTION 5166012 Revision 1 Documentation. To view the most recent appropriate and detailed documentation, click to go to MR Service Documentation.