GE Healthcare · MRSignaHiSpeedExcite22257042225697

No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Tabl…

PDU was in stop mode.

Symptom

No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.

Error message

2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.

2225697 MDS link failure. The IPG failed to receive an MDS transmission echo on a second try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0012 Transmitted byte: 0x11 2225697 MDS link failure. The IPG failed to receive an MDS transmission echo on a second try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0012 Transmitted byte: 0x11

Applies to

Models
Signa, HiSpeed, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.