Pular para o conteúdo principal
GE Healthcare · MRExciteHiSpeed224860222523182244588

Accidental Push of E-stop Button.

Symptom

Emergency Stop Button Hit, How to Recover and Start System

Error message

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

2252318 The Driver Module Has Detected a Power Supply Fault . Actual Voltage = 50.5 Limit Voltage = 425.0 2252318 The Driver Module Has Detected a Power Supply Fault . Actual Voltage = 50.5 Limit Voltage = 425.0

2244588 The GP is reporting SGA Power Supply Internal Wiring Fault or Internal Power Supply Undervoltage 2244588 "The GP is reporting SGA Power Supply Internal Wiring Fault\n\ or Internal Power Supply Undervoltage. Reseat the internal\n\ SGA-PS connectors. Else, replace the SGA Power Supply." 2244588 "The GP is reporting SGA Power Supply Internal Wiring Fault or Internal Power Supply Undervoltage. Reseat the internal SGA-PS connectors. Else, replace the SGA Power Supply."

Applies to

Models
Excite, HiSpeed
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required