GE Healthcare · MRExcite222381720000232222572002223619200002321200002348
The NIC card is causing communication errors to the MGD Chassis.
Symptom
Applications Software Will Not Boot.
Error message
2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 200002322 No processes could be shutdown, so exiting. 200002322 lx-mr3t hostMonitor ssProcStop.c 1412 No processes could be shutdown, so exiting. 2257200 Failed to initialize MrMail. Fri Jan 28 08:15:45 2005 Error: 2257200 Host: lx-mr Process: WavRec File: WavRecMain.cc Line: 63 Failed to initialize MrMail. 2223619 RPC call failed. The NMRID is ifcc. 2223619 lx-mr3t ifcc mrmailmgr.c 501 RPC call failed. The NMRID is:ifcc. 200002321 No processes could be shutdown, skipping precursor limitations and trying again. 200002321 4 lx-mr3t hostMonitor ssProcStop.c 1399 No processes could be shutdown, skipping precursor limitations and trying again. 200002348 The System Software has terminated. Mon Mar 15 07:26:41 2004 Error: 200002348 Host: MMCLX_2 Process: hostMonitor File: ssProcStop.c Line: 1531 The System Software has terminated. 2223619 RPC call failed. The NMRID is:WavRec. 2223619 lx-mr3t WavRec mrmailmgr.c 501 RPC call failed. The NMRID is:WavRec. 200002352 A HOST_SHUTDOWN was initiated by the de
Applies to
- Models
- Excite
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Component
- SecondOpinion HP xw8000 PC 2368697-8 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.