Unable to Scan with TR bias Body open circuit Error after Mechanical Upgrade from 9X to HDX Performed by Mechanical Inst…
Unable to Scan with TR bias Body open circuit Error after Mechanical Upgrade from 9X to HDX Performed by Mechanical Installer.
Symptom
Unable to Scan with TR Driver Fault Detected Open Circuit. Source of Fault is Body TR Error after Mechanical Upgrade from 9X to HDX Performed by Mechanical Installer Upgrade from 9X to HDX. Unable to scan with "TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR" which began immediately after mechanical upgrade performed by Mechanical Installer. See Long Description for the complete error message string logged when a scan is attempted. PQR 13093664 Mon Dec 18 15:57:19 2006 2252293 mrs3 NSP SCP:DCB 2915 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: BODY Scan: 50332/1 The Driver Module Has Detected a Fault. Details of the Fault can be found in other error messages.-> Mon Dec 18 15:57:19 2006 2252357 mrs3 NSP APS:aps_cf 243 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: BODY Scan: 50332/1 Driver Module error occurred. Mon Dec 18 15:57:20 2006 2224666 1 mrs3 psc PscMainControl.cc 1354 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, possible data acquisition error. Try again or use Manual Prescan. Mon Dec 18 15:57:20 2006 2252305 mrs3 NSP SCP:DCB 1289 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: BODY Scan: 50332/1 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 10 mA Limit: 2400 mA->
Error message
2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred. 2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR. Thu Aug 5 11:00:38 2004 Error: 2252305 Host: fmri3te Process: NSP File: SCP:DCB Line: 1197 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 16 mA Limit: 2400 mA-> 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2319986 PLUG, 7/8 EIA RIGHT ANGLE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.