OrthoMRI Hard drive has bad sector.
Symptom
OrthoMRI System will not boot up.
Error message
OrthoMRI Disk boot failure.
Applies to
- Product
- Orthopedics E-Scan (Win) XQ. Orthopedics E-Scan (Win) XQ.
- Software
- OrthoMRI 9.1b revision software.
- Component
- OrthoMRI Hard Drive for E-scan XQ, p/n LNR9102062004. See Long Description Note: Always use the most current revision of service documentation. When you install a new hard drive preloaded with windows 9.1b software it doesn't have the DST patch installed. You will need to install the patch. The patch cannot be ordered separately if you don't have one. You must order software kit LNR9101137066 from Coakley-tech. The kit includes the following: 9.1B Software – LNR9101137066 CD 9.0 E-Scan User manual LNR8300234308 CD 9.0 C-Scan User manual LNR8300198301 CD Daylight Savings Time patch LNR 9101137075CD Daylight savings patch instructions LNR 9101137075-LIT
- Area
- MR