Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2244647

Will Not Scan with Gradient Errors Due to Gradient Switch.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244647 The GSC reported that the gradient axis was in the wrong position Wed Feb 23 14:25:03 2005 Error: 2244647 Host: PMH-MR2 Process: NSP File: SCP:GP Line: 700 The GSC reported that the gradient Z axis was in the wrong position. If recovery is not successful, please try a TPS reset.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
Area
MR