GE Healthcare · MRSignaOvationExcite2237278
Fails TPS Reset when booting up to Applications or when Resetting TPS.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2237278 CERD BOOT ERROR - UNEXPECTED RESPONSE FROM CAP. Error: Wed Apr 27 19:20:12 2005 Error: 2237278 Host: MR4 Process: NSP File: TPS:cx_dpr_utils.c Line: 131 CERD BOOT ERROR - UNEXPECTED RESPONSE FROM CAP; The CAP's response to a TPS CERD Boot command is unknown.A valid response was not received for command 36 from TPS to CAP. unknown response: -1
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.