Defective 8-port Ethernet Switch in the System Cabinet.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2249420 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Error: 2249420 Host: MR01 Process: diagCoord File: excLog.c Line: 1392 Task ID: 0x3dcdef0 Time: 47275623 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Additional information in /usr/g/service/log/mxExLg-2d15c34

2248199 APS IrfDriver RRF link tx not ready error 7352 Thu May 19 10:38:33 2005 Error: 2248199 Host: pzmrov Process: NSP File: APS:IrfDriver::irfInit Line: 5665 IrfDriver RRF link tx not ready error 7352 Please perform a TPS Reset.

2248267 The RHO_DDS freq was set to 0 Hz. This value is too small. 2248267 11X, E2, G3,OPENMR4,5 error code is changing to 2248293 Post 11.1, E2M4a, G3M4b, OPENMR4, 5 sw 1121284117 6 1 Wed Jul 13 19:48:37 2005 2248267 4 lx-mr NSP SCP:cex_set_center_freq 1207 The RHO_DDS freq was set to 0 Hz. This value is too small. RHO_DDS will be set back to its default value. Probably a software error.

2223846 Download status packet from APS is not received by NSP 2223846 NSP TardisReset.cc 1181 Download status packet from APS is not received by NSP.

2248711 IrfDriver initialization failed 1071085863 2 1 Wed Dec 10 19:51:03 2003 2248711 4 MRS1 N

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.