GE Healthcare · MRExcite2249420472756232248199224826722482931121284117
Defective 8-port Ethernet Switch in the System Cabinet.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2249420 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Error: 2249420 Host: MR01 Process: diagCoord File: excLog.c Line: 1392 Task ID: 0x3dcdef0 Time: 47275623 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Additional information in /usr/g/service/log/mxExLg-2d15c34 2248199 APS IrfDriver RRF link tx not ready error 7352 Thu May 19 10:38:33 2005 Error: 2248199 Host: pzmrov Process: NSP File: APS:IrfDriver::irfInit Line: 5665 IrfDriver RRF link tx not ready error 7352 Please perform a TPS Reset. 2248267 The RHO_DDS freq was set to 0 Hz. This value is too small. 2248267 11X, E2, G3,OPENMR4,5 error code is changing to 2248293 Post 11.1, E2M4a, G3M4b, OPENMR4, 5 sw 1121284117 6 1 Wed Jul 13 19:48:37 2005 2248267 4 lx-mr NSP SCP:cex_set_center_freq 1207 The RHO_DDS freq was set to 0 Hz. This value is too small. RHO_DDS will be set back to its default value. Probably a software error. 2223846 Download status packet from APS is not received by NSP 2223846 NSP TardisReset.cc 1181 Download status packet from APS is not received by NSP. 2248711 IrfDriver initialization failed 1071085863 2 1 Wed Dec 10 19:51:03 2003 2248711 4 MRS1 N
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.