Erbtec 84 Fault Reported on Powerup.
Symptom
Erbtec 84 Fault Reported on Powerup. Unable to Scan. RF, MDS Link & SSM-Related Errors Scrolling in Message Log with System Idle. All cabinets and modules are powered on. Cycling system power and rebooting the system is no help. In checking the message log we found that RF, MDS Link & SSM-related errors were continually scrolling in the message log with the system idle. See Long Description for a sample of the errors scrolling over and over in the message log. Wed Mar 21 12:24:09 2007 2225513 1 lx-mr NSP HPC_PORT_ID:spi_tr_monitor 6255 Attention: T/R driver in service, not product mode. EN 165 SR 178 1174479849 1 1 Wed Mar 21 12:24:09 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 178 1174479850 0 1 Wed Mar 21 12:24:10 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 359 1174479850 1 1 Wed Mar 21 12:24:10 2007 2225663 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3060 ERBTEC RF Amplifier fault 84. RF power monitor fault. Proper signals are not being received from RF power monitor. Possible defective power monitor, cables, Processor Board, or power monitor microprocessor power momentarily dropped causing a reset. EN 359 SR 342 1174479862 0 1 Wed Mar 21 12:24:22 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 351 1174479862 1 1 Wed Mar 21 12:24:22 2007 2225699 1 lx-mr NSP HPC_PORT_ID:spi_mds 509 MDS link failure. An MDS device failed to respond to IPG on a 2nd try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 506 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cab) MDS opcode: 0x0A EN 351 SR 315 1174479864 0 1 Wed Mar 21 12:24:24 2007 2225550 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6300 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. Present State: 0x00 Requested State: 0xF0 EN 315 SR 176 1174479864 1 1 Wed Mar 21 12:24:24 2007 2225538 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6245 T/R driver indicates it is no longer in a normal mode of operation. EN 176 SR 178 1174479864 2 1 Wed Mar 21 12:24:24 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 342 1174479868 0 1 Wed Mar 21 12:24:28 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 351 1174479869 0 1 Wed Mar 21 12:24:29 2007 2225699 1 lx-mr NSP HPC_PORT_ID:spi_mds 509 MDS link failure. An MDS device failed to respond to IPG on a 2nd try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 507 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cab) MDS opcode: 0x0A EN 351 SR 353 1174479871 0 1 Wed Mar 21 12:24:31 2007 2225704 1 lx-mr NSP HPC_PORT_ID:spi_mds 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between cabinets. Run MDS fiber optic link diagnostics.-> EN 353 SR 176 1174479911 0 1 Wed Mar 21 12:25:11 2007 2225538 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6245 T/R driver indicates it is no longer in a normal mode of operation. EN 176 SR 315 1174479911 1 1 Wed Mar 21 12:25:11 2007 2225550 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6300 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. Present State: 0x00 Requested State: 0xF0 EN 315 SR 178 1174479911 2 1 Wed Mar 21 12:25:11 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 359 1174479911 3 1 Wed Mar 21 12:25:11 2007 2225663 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3060 ERBTEC RF Amplifier fault 84. RF power monitor fault. Proper signals are not being received from RF power monitor. Possible defective power monitor, cables, Processor Board, or power monitor microprocessor power momentarily dropped causing a reset. EN 359 Wed Mar 21 12:29:32 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 351 1174480172 1 1 Wed Mar 21 12:29:32 2007 2225699 1 lx-mr NSP HPC_PORT_ID:spi_mds 509 MDS link failure. An MDS device failed to respond to IPG on a 2nd try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 506 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cab) MDS opcode: 0x0A EN 351 SR 176 1174480174 0 1 Wed Mar 21 12:29:34 2007 2225538 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6245 T/R driver indicates it is no longer in a normal mode of operation. EN 176 SR 315 1174480174 1 1 Wed Mar 21 12:29:34 2007 2225550 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6300 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. Present State: 0x00 Requested State: 0xF0 EN 315 SR 178 1174480174 2 1 Wed Mar 21 12:29:34 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 342 1174480197 0 1 Wed Mar 21 12:29:57 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 351 1174480198 0 1 Wed Mar 21 12:29:58 2007 2225699 1 lx-mr NSP HPC_PORT_ID:spi_mds 509 MDS link failure. An MDS device failed to respond to IPG on a 2nd try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 506 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cab) MDS opcode: 0x0A EN 351 SR 176 1174480199 0 1 Wed Mar 21 12:29:59 2007 2225538 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6245 T/R driver indicates it is no longer in a normal mode of operation. EN 176 SR 315 1174480199 1 1 Wed Mar 21 12:29:59 2007 2225550 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6300 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. Present State: 0x00 Requested State: 0xF0 EN 315 SR 178 1174480199 2 1 Wed Mar 21 12:29:59 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 342 1174480200 0 1 Wed Mar 21 12:30:00 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 310 1174480200 1 1 Wed Mar 21 12:30:00 2007 2225696 1 lx-mr NSP HPC_PORT_ID:spi_mds 506 MDS link failure. A receive error was detected on the echo of an MDS retry transmission. Check the connections and signal level on the MDS fiber optic link. Timer: 254 UART status: 0xEB Received byte: 0x13 EN 310 SR 178 1174480202 0 1 Wed Mar 21 12:30:02 2007 2225481 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. EN 178 SR 342 1174480256 0 1 Wed Mar 21 12:30:56 2007 2225693 1 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A EN 342 SR 351 1174480256 1 1 Wed Mar 21 12:30:56 2007 2225699 1 lx-mr NSP HPC_PORT_ID:spi_mds 509 MDS link failure. An MDS device failed to respond to IPG on a 2nd try. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 506 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cab) MDS opcode: 0x0A EN 351 SR 176 1174480257 0 1 Wed Mar 21 12:30:57 2007 2225538 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6245 T/R driver indicates it is no longer in a normal mode of operation. EN 176 SR 315 1174480257 1 1 Wed Mar 21 12:30:57 2007 2225550 1 lx-mr NSP HPC_PORT_ID:spi_tr_mon 6300 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. Present State: 0x00 Requested State: 0xF0
Error message
2225663 ERBTEC RF Amplifier fault 84. Power Monitor Fault. 2225693 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B 2225693 lx-mr NSP HPC_PORT_ID:spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 254 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0B 2225481 RF Amplifier Error. The RF amplifier is in factory, not product mode. 2225481 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3055 RF Amplifier Error. The RF amplifier is in factory, not product mode. 2225550 Multi-coil array driver error. A write to the multi-coil array driver coil register failed to verify See Long Description 2225550 Host: LX_MR1 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6300 Multi-coil array driver error.^A A write to the multi-coil array driver coil register failed to verify. Possible causes: bad board or connections in the TR chassis. 2225699 MDS link failure. An MDS device failed t
Applies to
- Models
- Signa, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.