Pular para o conteúdo principal
GE Healthcare · X-ray R&FAdvantx

Advantx Systems with line Regulator part number: 46-232584G1-C or older.

Advantx power up problems.

Symptom

Advantx Systems with line Regulator part number: 46-232584G1-C or older. When the Advantx powers down, the Line Regulator Capacitors C19 and C20 hold up the DC Supply while the line drops out. This produces an error signal that causes the 120vac line regulator to drive up by 20-30 volts. When the unit is next turned on, it applies an excessively high voltage to all modules and components down line causing damage. Sometimes the regulator will drive down and if the voltage is too low, boot and power up problems occur.

Applies to

Models
Advantx
Product
Advantx
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray R&F issues