Advantx 1 - Improper calibration data entered during calibration of Advantx Cal Unit P314 "SID Receptor XX Hor…
Advantx 1 Compax 40E systems with SG100 Wall Receptor Cal unit P314.
Symptom
Advantx 1 - Improper calibration data entered during calibration of Advantx Cal Unit P314 "SID Receptor XX Horizontal" resulting in poor collimator to receptor sizing. Advantx 1 systems configured with Compax 40E/SG100 require cassette tray to be inserted in the receptor for proper SID feedback data to be stored in the database when exiting unit. Calibration users guide does not mention this requirement which is unique to this system configuration.