Pular para o conteúdo principal
GE Healthcare · MRExcite2252301

Unable to Scan with the 1.5T 8Ch TR Knee Array due to Multicoil Bias Fault Detected - Open Circuit Always on Channel 6.

Symptom

Unable to Scan with the "1.5T 8Ch TR Knee Array" due to "Multicoil Bias Fault Detected - Open Circuit" Always on Channel 6 See Long Description Unable to scan with the "1.5T 8Ch TR Knee Array" due to "Multicoil Bias Fault Detected - Open Circuit", always on Channel 6. (See below for a copy of the logged error from the site). This doesn't happen with any other coil, and it occurs with the coil plugged into either the "A" or "B" Excite HD LPCA coil ports. Mon Feb 19 14:38:07 2007 2252301 1 MR01 NSP SCP:DCB 918 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD2 Channel 6 of 16, counting from 1 Value: 3 mA Limit: 98 mA Mask of failed channels: 0x20 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B->

Error message

2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicated by the most significant bit. For example, a mask of 0x3c converts to 0011 1100, meaning channels 3, 4, 5, and 6 all experienced the same failure. The key at the bottom of the error message shows the MC driver bias channels provided by the Driver Module used with each mult

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion 1.5T 8 Channel TR Knee Array 5114356-2 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues