System Not Recognizing L_BREAST, R_BREAST, and 2_BREAST with Coil MRID 1.5/1.0T Open Breast Coil M1087BT MRID coil p/n 2…
System Not Recognizing L_BREAST, R_BREAST, and 2_BREAST with Coil MRID 1.5/1.0T Open Breast Coil M1087BT MRID coil p/n 2246360 Connected.
Symptom
System Not Recognizing L_BREAST, R_BREAST, and 2_BREAST with Coil MRID 1.5/1.0T Open Breast Coil M1087BT p/n 2246360 Connected. The Coil ID LED Stays Dark or Off. See Long Description System not recognizing "L_BREAST", "R_BREAST", and "2_BREAST" which are the defined coil names for the MRID 1.5/1.0T Open Breast Coil M1087BT, coil p/n 2246360 with this coil connected. The "Coil ID" indicator LED stays dark, it does not illuminate. There are no errors logged, but the FE stated that the operator sees a popup message to the effect that the coil is not recognized, but he doesn't have the exact text of that popup message. He can scan a phantom okay with this coil using the "geservice" patient ID which causes the system to ignore Coil ID. The customer tried a different protocol but there was no change to symptoms. All other coils are working fine.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2246360-3 Coil ID Cable This Coil ID cable is for the MRID 1.5/1.0T Open Breast Coil M1087BT coil This Coil ID cable is for the MRID 1.5/1.0T Open Breast Coil M1087BT coil
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.