GE Healthcare · MR
During LCC 300 Magnet Early Readiness Training, the gauss limits for magnet service was misstated as having no…
Documentation is Wrong
Symptom
During LCC 300 Magnet Early Readiness Training, the gauss limits for magnet service was misstated as having no ferromagnetic tools, equipment or components inside the 5 gauss zone of a ramped magnet MR Smile 4/13/2004: Gauss Limits For Magnet Service Effectivity: All GE Superconducting Magnets
Applies to
- Product
- Magnet LCC 300. Software Version Concept should NOT be selected for Magnet Products Low Cryogen Consumption Serial number:Wxxx Field strength: 3.0 T He Vessel Pressure:>3 psig but <5 psig Shield Temp:< 55 Kelvin Coldhead Temp:> 3.8K but < 4.5K Recondenser Temp:> 4.2K but < 4.55K Delta Temp (Coldhead - Recondensor):> 0 but < 0.3K Min. LHe Fill Level: 60% Min. LHe Ramp Level: 100% Min. LHe Shim Level: 80% Helium Volume:1735 Liters Burst Disk:2349529 Burst Disk O-ring:46-281101P1 Baffle Assy: 2180818 Shim Lead Large 5/8:2173354
- Area
- MR
Full cause & step-by-step resolution