Advantx - To inform field installation personel that the wire colors for MIS10247A-D are shown incorrectly in…

Advantx Wire Colors For MIS 10247B On Advantx Rad Systems Service Note 3329.

Symptom

Advantx - To inform field installation personel that the wire colors for MIS10247A-D are shown incorrectly in the MIS Chart manuals. Advantx RAD Systems must NOT follow the re-wiring diagram of connector A1 A21 A1 J102 as shown on MIS chart 10247B in Direction 46-019338 Rev.7, Advantx RAD System MIS Charts. Although the MIS charts for MIS 10247A&D (MVP & MVP Micro) and MIS 10247C (Advantx R&F) show the incorrect colors, there should be no affect since no re-wiring is performed on these systems. All the affected MIS charts will be corrected in the next revision of the MIS Chart directions.

Applies to

Models
Advantx
Product
Advantx
Area
X-ray R&F