Advantx - There is a performance problem that might occur on early production units of the Advantx Console Tou…

Advantx Console Touch Screen, Ambient Light Level Problem.

Symptom

Advantx - There is a performance problem that might occur on early production units of the Advantx Console Touch Screen Frame. This problem is typically seen when the ambient light conditions are changed after the Console is powered-up. Under these conditions, the Console may appear to lock-up, with no system response when the display is touched. (To confirm that this is the problem, reset the Console CPU Board when it stops responding to touch input. This will force the touch frame to re-calibrate itself for the new Ambient Light conditions.)