Advantx-E Error 8541: Temperature out of range was found in the system error log and the logged temperature wa…
Advantx-E Error 8541 Temperature out of range.
Symptom
Advantx-E Error 8541: Temperature out of range was found in the system error log and the logged temperature was 0 degree C. In the current Titan software design and implementation, the temperature monitoring background task will access the Titan temperature monitoring hardware periodically to read temperature data. Under normal operation environment Error 8541 indicates potential Titan System temperature monitoring sensor hardware failure or system operation environment temperature out of range.
Error message
Error 8541Temperature out of range
Applies to
- Models
- Advantx
- Product
- Advantx-E
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- ADXE RFX/SFX Legacy popping noises during angulation SN 3715
- Advantx Rad Compax 40E system with SG100 Tilting Vertical Bucky Stand.
- Advantx Legacy system with DRS 3.2 with DRS Gateway.
- Digital Rad Error 6D00 on Advantx-E console.
- Advantx Erratic shutter operation Adjust motor and gear for proper mesh