GE Healthcare · MRExciteSigna221853722481212248131224810210151

Can Not Scan. Replace DRF Board.

Symptom

Can Not Scan

Error message

2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248

2248121 APS Send DRF 4 cmd Message Error (no response)

2248131 APS DC Send DRF info error -1 Tue May 4 20:27:03 2004 Error: 2248131 Host: lx-mr Process: NSP File: APS:DacqControl::processPkt Line: 536 APS DC send DRF info error -1 Software error, please perform a TPS Reset.

2248102 Received Invalid cm Packet 10151, scan state = 3 Tue May 4 20:27:03 2004 Error: 2248102 Host: lx-mr Process: NSP File: APS:processPkt Line: 896 Received invalid cm packet 10151, scan state = 3 Software error, please perform a TPS Reset.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR