Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova50014514923351198525119853

Gantry moving slow due to DETECTOR collision sensor errors.

Symptom

Gantry moving slow due to DETECTOR collision sensor errors. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. Gantry moving slow due to detector collision errors. This issue is caused by a collision detected on the Detector side of the gantry. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. System impact: Message displayed on monitors and BIP on TSSC every 30 seconds. Full gantry motion. Displayed message : "Collision detection failure, continue with care; Otherwise call Service" The Hardware code tells whether the problem is located on the Detector or on the Tube side: HW=4 for the Detector Faceplate HW=8 is the Detector Bumper and HW=10 for the Tube . Check the anti collision sensors on the Detector. Touch the collision sensor margin on one side to re-balance the collision sensors. Touch the collision sensor margin on both sides to clear the collision error. If the problem persists (i.e., the error is not cleared), replace the corresponding ANTI-COLLISION COVER

Error message

50014, error Text = POS : II COLLISION, debug Text /= pos-error=4A12, hw-status=0008, 2nd level bump System impact: Half speed until system level1 collision is cleared. Error message cleared following 2 seconds. Displayed message : "Detector collision: reverse motion to continue". A collision is detected on the bladder 4A12 with Hardware 0008 indicates a 2nd level bump, not a level 1 bump as indicated by this concept Bolus System, Install the Bolus M4+ (kit #5149233) on the system. BOLUS KIT 1 and 2 to replace SW package . 5119852 rev 7 BOLUS UPGRADE KIT 1 for INNOVA 4100 3100. 5119853 rev 7 BOLUS UPGRADE KIT 2 for INNOVA 4100 3100. 50014 WR VAS / POS POS : II COLLISION Debug: hw-status=4 pos-error=4a12 usr-status=0 POS ERROR=4A12,HDW STATUS=0004,USR STATUS=0000

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required