Pular para o conteúdo principal
GE Healthcare · MRExcite2220002

Unable to Scan Patients Unless Patient ID is GEservice.

Symptom

Unable to Scan Patients Unless Patient ID is GEservice New Excite HD system which has successfully generated the IIS (Install-In-Spec) key is unable to scan patients due to error message (See Error Concept for error message). Note that system will scan provided "geservice" is used in the Patient ID field but of course this is not permissible when scanning patients.

Error message

2220002 System calibration is not complete. The system is not ready for diagnostic imaging.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution