Advantx VIC - Setting G3 Voltage on New A9 Board When changing the VIC A9 power supply board, the label on the…
Advantx VIC - Setting G3 Voltage on New A9 Board
Symptom
Advantx VIC - Setting G3 Voltage on New A9 Board When changing the VIC A9 power supply board, the label on the back of the board can be misleading. The label on the old board, part number 46-321182G1-C states that G3 SHOULD BE SET TO 550 VOLTS FOR ALL SYSTEMS EXCEPT SYSTEMS WITH DX 1024. SET TO 700 VOLTS FOR DX 1024. The label on the new board, part number 46-321182G2-D states that G3 SHOULD BE SET TO 550 VOLTS FOR ALL SYSTEMS.
Applies to
- Models
- Advantx
- Product
- Advantx
- Component
- VIC Camera Video Image Controller
- Area
- X-ray R&F
Full cause & step-by-step resolution