VIC - SN 3777 Camera Lens Leveling Problem

Symptom

VIC - Camera Lens Leveling Problem Proper leveling of camera lens cannot be performed in some systems. The magnetic plate 2164221 mounted to the front of the camera head lacks end slots in 20-30 units shipped. The plate is nickel plated to block electrical fields, and is made of mu metal to block magnetic fields. Magnetic fields found on tilt-c systems with DLX induse noise on subtraction images that appears similar to noise caused by microphonics.

Applies to

Models
Advantx
Product
Advantx
Component
VIC Camera Video Image Controller
Area
X-ray R&F