GE Healthcare · X-ray R&FAdvantx
VIC Camera Head H-Sweep Board - Sweep size and center off.
Symptom
VIC Camera Head H-Sweep Board - Sweep size and center off. Diagnostic Error Code: 8762 Horizontal size and center were both properly in adjustment on 8/1/97. Currently the image is approximately twice the correct size and shifted to the right by about 50% as viewed on the monitor. FE was able to adjust the conventional image back into spec but he was at one limit of the adjustment. The digital images are so far out that they cannot be adjusted.
Error message
Error 8762 Error 8762
Applies to
- Models
- Advantx
- Product
- Advantx
- Component
- VIC Camera Video Image Controller
- Area
- X-ray R&F
Full cause & step-by-step resolution