Intermittent Omega V table longitudinal feedback pot

Symptom

Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211 Bolus run Aborted Failure ID ADXR KO: Bolus Chasing 1.0 Focal Spot Comment: Xrays Aborted after xx Images by error 50211

Error message

50118 POS Table: Encoder, potentiometer failure; Debug: , HDW STATUS=0000 or HW 0001,USR STATUS=XXXX followed by 50211 POS Table: Motion control error during Bolus procedure. Both occur at the same time during a Bolus run. 50118 POS Table: Encoder, potentiometer failure; Debug: POS ERROR=4AC3, HDW STATUS=0000,USR STATUS=XXXX 50211 POS Table: Motion control error during Bolus procedure. Debug: [XRImChainPos::logPosError] Reporing the Motion Control error for GantryMotionControlError See LD for complete error. 50118 POS Table: Encoder, potentiometer failure; Debug: POS ERROR=4AC3,HDW STATUS=0000,USR STATUS=XXXX. HW and User status indicate Longitudinal Feedback pot problem. - Longitudinal Feedback pot problem 50118 POS Table: Encoder, potentiometer failure; Debug: , HDW STATUS=0001,USR STATUS=XXXX followed by 50211 POS Table: Motion control error during Bolus procedure. Both occur at the same time during a Bolus run. This indicates a problem with the Lateral Table Feedback Pot During a bolus run. Recalibrate the Table Lateral Potentiometer P403 Check the alignment and correct any mounting hardware issues 50211 POS Table: Motion control error during Bolus procedure. Debug: [XRImChainP

Applies to

Models
Innova, Advantx
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
Bolus M3 DL release : DL_LBYBOLUSM3_6.17.050317; JEDI release : B6.30BFB. Ghost DL-3 : 18.7 Ghost DL-2 : 18.5 Ghost DL-5 : 20.4 DL : DL_LBYBOLUSM3_6.17.050317 DL Atlas Core : 3.1.R RTAC Atlas Core : 3.1.R IDC : 4.5.4 Calypso : 2.19 Calypso Uptime : BOLUS1.17.1 Positioner : 1.2.24B_IS6.5 Jedi : B6.30BFB
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.