Pular para o conteúdo principal
GE Healthcare · MRExcite2252239225236070727

SCP did not respond to the start packet of WavRec and Startup Timeout was detected on the Drivers Control Board Due to t…

SCP did not respond to the start packet of WavRec and Startup Timeout was detected on the Drivers Control Board Due to the Driver Board

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2252239 Startup Timeout was detected on the Drivers Control Board

2252360 The Driver Module did not detect the T/R DD board, which is required Error: 2252360 Host: MR03 Process: diagCoord File: CanDriverHWO Line: 2721 Diagnostic: DCB Diag Iteration: 1 of 1 Action: 1 Task ID: 0x3156958 Time: 70727 Object: SCP T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset.-> Extended Error: 2252360 ERROR: EM_SCP_TRDD_NOT_PRESENT T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset. Possible cause(s): Driver Module

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.