GE Healthcare · MRExcite2252239225236070727
SCP did not respond to the start packet of WavRec and Startup Timeout was detected on the Drivers Control Board Due to t…
SCP did not respond to the start packet of WavRec and Startup Timeout was detected on the Drivers Control Board Due to the Driver Board
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2252239 Startup Timeout was detected on the Drivers Control Board 2252360 The Driver Module did not detect the T/R DD board, which is required Error: 2252360 Host: MR03 Process: diagCoord File: CanDriverHWO Line: 2721 Diagnostic: DCB Diag Iteration: 1 of 1 Action: 1 Task ID: 0x3156958 Time: 70727 Object: SCP T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset.-> Extended Error: 2252360 ERROR: EM_SCP_TRDD_NOT_PRESENT T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset. Possible cause(s): Driver Module
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.