Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova21201188228000592800060280006128000662800006

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset.

Innova System boot failure

Symptom

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset. Customer needs to boot again to recover. When Innova system has a actor issue.

Error message

RTAC displays OCP Code 69H on Boot. 69h = Setup System Management Mode (SMM) area RTAC OCP Code of 69 on Boot. 69h = Setup System Management Mode (SMM) area The OCP code maybe seen as b9 (B9) 212011882 05/11/2010 13:02 GMT

2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout 2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout. May be followed by error codes 2800060, 2800061, 2800066, 2800006 2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout May be followed by error codes 2800060, 2800061, 2800066, 2800006

2800060 IH XRSubSysMgrDL Failure of XRDataMgr during boot of the application Debug:component startup timeout 2800060 IH XRSubSysMgrDL Failure of XRDataMgr during boot of the application Debug: component startup timeout

2800061 IH XRSubSysMgrDL Failure of XRUserIfDL during boot of the application Debug: component startup timeout

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Component
RTAC in Innova RTAC in Innova RTAC in Innova Rtac's 128 meg memory stick 128MB Memory stick? Option 2 of the SN says to order FMI 12044 pn 5160221-3 if 128MB memory. cpu + memory from RTAC 1 chassis, part number 2271482 and swap memory the 128 meg memory stick into the old chassis. 128MB Memory stick? Option 2 of the SN says to order FMI 12044 pn 5160221-3 if 128MB memory cpu + memory from RTAC 1 chassis, part number 2271482 and swap memory the 128 meg memory stick into the old chassis.
Area
X-ray Vascular

Full cause & step-by-step resolution