Propeller Scan Option Will Not Run.

Symptom

Propeller Scan Option Will Not Run or Runs Very Slow propeller option will not function

Error message

2257001 This application cannot be run on this system due to insufficient image reconstruction power.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2325947-13 AP Reflex Minerva II 400B Board Slave Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave