MC Power Supply output too low

Symptom

Unable to scan Phased Array Shoulder Coil due to PA Bias Short errors on Profile5. Getting PA Bias 2262724 errors intermittently in APS on site PA Shoulder protocol 3-plane loc. Going into MPS and adjusting TG allows the scan to continue. When TG=60, failure occurs. When TG adjusted manually in MPS to 90-126, scans ok. Happens on PA Shoulder large & small cups.

Error message

2262701 Pre amplifier is not ready. 2262701 NSP APS:aps_pass1 276 Pre amplifier is not ready.

2262724 Phased Array bias error(short) is occurred 2262724 NSP SCP:CoilIdSrpHWO 538 Phased Array bias error(short) is occurred.

Applies to

Models
Excite
Product
Profile with Excite
Software
Profile5 Applies to Profile5 (Excite) only. Profile5 is 12.x based software, not ProfileHD.