Pular para o conteúdo principal
GE Healthcare · MRExcite22279942218527

Insufficient TPS memory installed to support large data collection pulse sequences back-to-back.

Symptom

Unable to Scan with Not Enough or Insufficient TPS memory available Errors See Long Description Trying to set up for another scan and the system errors stating the reference scan stopped Insufficient TPS memory. 2218528 Host: lx-mr Process: scn File: MsaScanOpsScrn.m Line: 1505 Reference Scan has stopped. Insufficient TPS memory available.

Error message

2227994 The memory or other resources may be insufficient. Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wai Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wait

2218527 Scanning has stopped. not enough memory available to start acquisition, please try again. 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again. Error: 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution