GE Healthcare · MRExcite22279942218527
Power cycling can corrupt TPS Communication and Operating System software.
Symptom
Do not power cycle GOC or System Cabinet on 14x without shutting down the ICN's One cannot cycle power any longer to the GOC or the System Cabinet. There are 2 ICN's (Image Compute Nodes) in the VRE which have operating software downloaded from the Host. They must be shutdown prior to cycling power to the GOC or System Cabinet otherwise the OS software can become corrupt.
Error message
2227994 The memory or other resources may be insufficient. Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wai Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wait 2218527 Scanning has stopped. Insufficient TPS memory available. See Long Description Mon Feb 5 11:06:05 2007 Error: 2218527 Host: mr3t1 Process: scn File: MsaScanOpsScrn.m Line: 1036 Scanning has stopped. Insufficient TPS memory available.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.