GE Healthcare · MRExcite22279942218527

Power cycling can corrupt TPS Communication and Operating System software.

Symptom

Do not power cycle GOC or System Cabinet on 14x without shutting down the ICN's One cannot cycle power any longer to the GOC or the System Cabinet. There are 2 ICN's (Image Compute Nodes) in the VRE which have operating software downloaded from the Host. They must be shutdown prior to cycling power to the GOC or System Cabinet otherwise the OS software can become corrupt.

Error message

2227994 The memory or other resources may be insufficient. Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wai Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wait

2218527 Scanning has stopped. Insufficient TPS memory available. See Long Description Mon Feb 5 11:06:05 2007 Error: 2218527 Host: mr3t1 Process: scn File: MsaScanOpsScrn.m Line: 1036 Scanning has stopped. Insufficient TPS memory available.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution