TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx. 3T Excite HDx 14.0. 2261238 The ASC BB IF Board was d…
TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx. 3T Excite HDx 14.0. 2261238 The ASC BB IF Board was detected but is not configured in MRconfig.cfg
Symptom
TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx After Installing the CPC MNS optional RF amplifier, and configuring the software for the correct MNS amplifier, the TPS fails reset with config errors.
Error message
2261238 The ASC BB IF Board was detected but is not configured in MRconfig.cfg. 2261238; The ASC BB IF Board was detected but is not configured in MRconfig.cfg. Set usssType=yes in the MRconfig.cfg file, verify that the RF Amp cabling is connected to the ASC BB IF Board, and perform a TPS Reset
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.