TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx. 3T Excite HDx 14.0. 2261238 The ASC BB IF Board was d…

TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx. 3T Excite HDx 14.0. 2261238 The ASC BB IF Board was detected but is not configured in MRconfig.cfg

Symptom

TPS Fails Reset After CPC MNS Amp Option Installation on a 3T HDx After Installing the CPC MNS optional RF amplifier, and configuring the software for the correct MNS amplifier, the TPS fails reset with config errors.

Error message

2261238 The ASC BB IF Board was detected but is not configured in MRconfig.cfg. 2261238; The ASC BB IF Board was detected but is not configured in MRconfig.cfg. Set usssType=yes in the MRconfig.cfg file, verify that the RF Amp cabling is connected to the ASC BB IF Board, and perform a TPS Reset

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution