GE Healthcare · MRExcite22564582256853224942047275623
Intermittently unable to download or scan the Propeller PSD Due to the Eight Port Ethernet Switch.
Symptom
Unable to Download or scan the Propeller PSD Unable to download or scan the Propeller PSD unless the TPS has been reset. Nearly every Propeller scan fails. Sometimes unable to scan other PSDs after a Propeller scan has been run, until the TPS has been reset.
Error message
2256458 Server switch failed! 2256458 lx-mr LxMxConsumer LxMxSwitcher.cpp 292 Server Name: SC Server switch failed! 2256853 The ACE Timer monitor timed out. 2256853 lx-mr LxMxConsumer MonitorHost.cpp 366 Server Name: SC The ACE Timer monitor timed out. 2249420 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Error: 2249420 Host: MR01 Process: diagCoord File: excLog.c Line: 1392 Task ID: 0x3dcdef0 Time: 47275623 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Additional information in /usr/g/service/log/mxExLg-2d15c34
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.