GE Healthcare · X-ray VascularInnova
When using PST cannot download software to MCB (motor control) boards in positioner.
Symptom
When using PST cannot download software to MCB (motor control) boards in positioner. When using PST cannot download software to MCB boards in positioner.
Error message
After 72% on process bar the system displays "Timeout detected. Please check the port or serial cable." in the Firmware Download Application screen.
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- CPU Roadunner in the Positioner CPU Roadunner in the Positioner. Check FeTools and ASM for any updates on part numbers. Aways load SW on new board. Part #: 5145126 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RP CPU ROADRUNNER CVC FRU: Yes Status: Orderable CPU Roadunner in the Positioner. Check FeTools and ASM for any updates on part numbers. When replacing CPU, Aways load SW on new board. Part #: 5145126 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RP CPU ROADRUNNER CVC FRU: Yes Status: Orderable
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.