Pular para o conteúdo principal
GE Healthcare · MRExcite224832122483552248346

TPS Download Fails with 2248321 2248346 and 2248355 Due to CAT 5 Cable from System cabinet Ethernet Switch to PC Etherne…

TPS Download Fails with 2248321 2248346 and 2248355 Due to CAT 5 Cable from System cabinet Ethernet Switch to PC Ethernet Card

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2248321 - The SC server is not responding, the server might be hung. Error: 2248321 Host: MR01 Process: NSP File: SCP:sct_wdog_exp Line: 1003 Communication Error. The callback for SCSM state 0x2 was not received in time. The SC server is not responding, the server might be hung. Please perform a TPS reset to recover.

2248355 AGP to SCP State Error. SCT received a wavegen status 3 in a non-wavegen state 3 It may be necessary to perform a TPS reset. Error: 2248355 Host: hopgemr1a Process: NSP File: SCP:sct_mct_wavegen_status Line: 741 AGP to SCP State Error. SCT received a wavegen status 3 in a non-wavegen state 3 It may be necessary to perform a TPS reset.

2248346 Pulse generation failed. Probable PSD software error. 2248346 MMCN NSP SCP:sct_prep_or_start 417 Pulse generation failed. Probable PSD software error. Pulse generation timed out. It may be necessary to perform a TPS Reset.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5160775 FRU ICN Ethernet Cable Collector This kit contains: 3 cables between ICN and Ethernet Switch, 1 J63 to Ethernet Switch cable, 1 Ethernet cable 100 ft (30.5m) in length between Host and J63 on System Cabinet, and 1 cable label book. HDx system.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues