Pular para o conteúdo principal
GE Healthcare · MRExcite2262937506332263078

Error 2263078 Software transmit overflow occurred and 2262937 Flash Programming Failed Due to UPM NB IF Board in the ASC…

Error 2263078 Software transmit overflow occurred and 2262937 Flash Programming Failed Due to UPM NB IF Board in the ASC Chassis

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2262937 FLASH Programming Failed Mon Dec 11 12:47:08 2006 Error: 2262937 Host: MR1 Process: NSP File: SCP:PiffNode Line: 802 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Spin Echo Coil: BODY Scan: 50633/1 PIFF Node RF NB Amp Interface: FLASH Programming Failed Possible bad FLASH device.

2263078 A software transmit overflow occurred on a UART link Mon Dec 11 12:47:09 2006 Error: 2263078 Host: MR1 Process: NSP File: SCP:../uartLib.c Line: 343 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Spin Echo Coil: BODY Scan: 50633/1 UART Communication Error A software transmit overflow occurred on a UART link. Source: RF NB Amp Interface UART: Debug UART

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
2391146 UPM NB AMP IF BOARD, also known as ASC RF Interface Board.