GE Healthcare · MRExcite
Elements mixed between 3T RF POWER MEASUREMENT KITs, invalidating their calibration.
Symptom
Elements mixed between 3T RF POWER MEASUREMENT KITs, invalidating their calibration. Three 3T RF POWER MEASUREMENT kits at the calibration vendor were found to have elements from other kits. The Bird power monitor meter and elements are calibrated as a kit. Mixing elements between kits invalidates the calibration of the instrument. See Resolution concept for the complete SMILe released 10/11/2006.
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution