Coil ID Only Works on One Side of the Table

Symptom

Coil ID Only Works on One Side of the Table See Long Description The HFO5 site can't get any coil to work on the left side of the MSHS2. When a coil is plugged into the left side of the table, the Coil Status LED turns red, indicating that the system senses a coil is plugged in there, but the system doesn’t recognize the coil. A popup message appears to the operator when starting a new exam that the coil is not recognized but no error message is logged. This happens for all coils. It always works on the right side of the table. Because of this problem the customer has been only using the right side of the table to plug in coils to be able to scan. We have swapped out the Head TR board in the MSHS2, the MSHS2 itself, the 8 channel switch, and the SRI3. Software has also been reloaded on the system, as it was suggested this might be a software problem. None of this fixed the problem.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Component
2384547 Cable Run # 1167 MG3-A4-J31 to PT1-J21
Area
MR